1、Apply for flush mount, measure and display environment T/RH such as cleanrooms in electronics or pharmaceutical plants, hospital and various laboratories, etc.
2、316L front panel, PMMA window, flat surface, no dust stays, good for all detergents, sanitizers and bactericides
3、H17 has large LCD, H17M has 3.2” TFT color LCD display
4、Special designed for some probe transmitters from well known brands such as E+E and Rotronic. High accuracy, 100% field interchangeable without re-calibration
5、2 channels 4~20mA output, with over voltage and reverse polarity protection
6、Strong EMI capability, meet the requirements of EN61326-1 industrial equipment
7、Compatible to any DDC/PLC/SCADA or other data collect and control systems
T/RH remote probe: (Refer to the operation manual from the original manufacturers for details)
Main unit and display:
(1) H17P: 4 bits LCD (with backlight), alternately display T/RH, display area 47x27mm, with figure H 16mm and unit H 2mm
(2) H17MP: 3.2" TFT color LCD (320×240), high light, can be set to display T/RH measurements and real-time curves, many display modes can be set by keys
Display panel material: PMMA
Resolution: 0.1°C(or °F, switchable), 0.1%RH
Update time: <1s
Power Supply:16~28VAC/19~35VDC
Consumption: H17P< 1.5 VA, H17MP< 2 VA
Analog outputs:2×4~20mA(3/4 wires), RL<500Ω; accuracy as low as 0.1%FS(T)/ 0.3%FS(RH)
Communication(H17MP):1xRS485/ModbusRTU,R/W,9600 baud rate; can set brightness and display setting functions, data transfer and calibration, etc. May set terminal resistance
Data(H17MP): Support data storage, trend curve display and data transfer
Keys(H17MP): can set display mode, parameter, brightness, unit and communication parameters, etc.
Terminals: max Ø1.5mm2
Work Environment: 0~50°C, 0~95%RH (no cond.)
Storage Environment: -10~70°C
Front panel material: 316L stainless steel, 1.5mm thick
Back housing parts: fire retardant ABS+PC(UL94V-0)
Protection: front panel IP65
Weight: H17P about 330g; H17MP about 380g
Approval: CE, meet EN61326-1 for industrial equipment
Note: For completeness, the above specifications include some of the probe's function and technical parameters. The products are combination of a main unit and a separated probe.
Models for the main unit (The product has two parts: the main unit and the probe. This is only for the main unit)

Models for the probe (The product has two parts: the main unit and the probe. This is only for the probe)

1. The cable should be ordered separately, with length of 2m, 5m or 10m. The models for E+E are E2, E5 and E10 respectively, and for Rotronic are R2, R5 and R10.
2. The main unit, probe and cable should be selected and ordered respectively. For example: the main unit is H17MP, the probe is EE07PFT3 and the cable is E5. The factory will deliver the three parts separately, and customer will install and connect them on site. Please refer to the operation manual for details.
3. In the environment where the pharmaceutical factory uses hydrogen peroxide disinfection, the hydrogen peroxide filter HA010115 is optional for the probe of E+E, and the hydrogen peroxide probe HC2A-S-HH is optional for Rotronic to resist the sensor erosion during the hydrogen peroxide disinfection process.
Display Modes for H17MP
Mode 1: T/RH display simultaneously with trend curve, shown as above
Mode 2: T/RH display simultaneously with trend curve Mode 3: T/RH display simultaneously


Mode 4: T/RH display alternately